The Telaire 7000 Series hand held monitors feature our patented absorption infrared technology.
Accurately measures CO2 and Temperature while calculating real-time ventilation rates. The Telaire 7000 Series also offers data logging and RH data gathering capabilities when purchased with available accessories. Product kits include a power adapter, however, the unit may be operated up to 80 hours on 4 AA batteries (not included).
Measurement Range: | 0-4,000 ppm voltage output 0-10,000 ppm display |
Pressure Dependence: | 0.13% of reading per mm Hg (Corrected via user input for elevation) |
Sensitivity: | ± 1 ppm |
Repeatability: | ± 20 ppm |
Temperature Dependence: | ± 0.1% of reading per °C or ± 2 ppm per °C, whichever is greater, referenced to 25°C |
Accuracy: | ± 50 ppm or 5% of reading, whichever is greater |
Annual Drift: | ± 20 ppm typical |
Response Time: | < 60 seconds for 90% of step change |
Warm-Up Time: | < 60 seconds at 22°C |
Operating Conditions: | 32 to 122° F (0-50°C) 0-95% RH, non-condensing |
Storage Temperatures: | -40 to 140°F (-40 to 60°C) |
Calibration Interval: | 12 months, offset adjustment using single gas at 0-1000 ppm CO2. Factory calibration available |
Temperature Channel:
Temperature Range | Voltage output 32 to 104°F (0 to 40°C) Display 32 to 122°F (0 to 50°C) |
Display Resolution | 0.1°F (0.1°C) |
Display Options | °F, °C, or Off. Set with panel button. |
Accuracy: | ± 2°F (± 1°C) |
Response Time: | 20-30 minutes (case must equilibrate with environment) |
Calibration Interval: | 12 months, offset adjustment using temperature standard at 50 to 86°F (10 to 30°C). Full factory calibration available |