HI84530 Titration SystemHI84530 Titration System

HI84530 HANNA Mini Titrator for Measuring Titratable Acidity

Availability: กรุณาสอบถามค่ะ

HI84530 HANNA Mini Titrator for Measuring Titratable Acidity in Water ใช้งานง่าย ทดสอบค่าความเป็นกรดในน้ำ
สอบถามเกี่ยวกับสินค้า
Product Tags

HI84530 HANNA

Mini Titrator for Measuring Titratable Acidity in Water

HI84530 HANNAThe HI84530 is an easy to use, fast and affordable, method specific titrator designed for testing titratable acidity levels in water. Based on an acid-base titration method, this mini titrator uses an optimized pre-programmed method of analysis with a powerful algorithm that determines the completion of the titration reaction potentiometrically with a glass body pH electrode.

The HI84530 incorporates a precision piston style dosing pump that adjusts the volume of dosing dynamically based on the voltage change. This type of dosing system reduces the amount of time required for the titration while providing for a highly accurate determination of the amount of titrant used.

This mini titrator is supplied complete with all the materials necessary to perform low and high range measurements of acidity in water. All chemicals are premixed and prepackaged including standardized titrants, reagents and pump calibration solution. There is no need for volumetric glassware or analytical balances.

Details

Acidity is an important parameter used to determine the quality of water (surface, drinking, wastewater). Acidity affects many things including chemical reaction rates, biological processes, and corrosiveness. It can also be used to monitor pollution in wastewater and drinking water.

Total titratable acidity is a quantitative measurement of the water sample’s ability to react with a strong base solution to a specific pH value. Many chemical species can contribute to the acidity measurement depending on the method of analysis. These species can include strong acids (hydrochloric acid, nitric acid, sulfuric acid, etc.), weak acids (acetic acid, carbonic acid, etc.), and hydrolyzing salts (iron, aluminum, manganese, etc.)

Double Junction pH Electrode

The HI84530 is supplied with the HI1131B refillable, double junction, combination pH electrode. By design, the HI1131B has a spherical tip for use in aqueous or liquid solutions. This versatile electrode provides a wide surface of contact with a sample and is ideal for direct measurements or acid base titrations in the water industry.

Piston Driven Dosing Pump

The heart of the HI84530 is the piston driven burette pump. This type of dosing system uses a motor in which each dose is very precisely controlled and the volume dispensed is accurately determined. The piston driven burette is controlled dynamically so that the volume of titrant being dosed is automatically adjusted based on a voltage response of the previous dose. This type of dosing speeds up the titration process by allowing for more titrant to be dosed at the start of the titration and then very small doses as the endpoint is reached.

Automatic Stirrer

The built-in stirrer is automatically maintained at a speed of 600 rpm, regardless of the viscosity of the solution being titrated.

Specifications

Range Low Range (as CaCO₃): 15.0 to 400.0 mg/L; 0.3 to 8.0 meq/L; High Range (as CaCO₃): 300 to 4000 mg/L; 6.0 to 80.0 meq/L

Resolution Low Range: 0.1 mg/L / 0.1 meq/L; High Range: 1 mg/L / 0.1 meq/L

Accuracy (@25ºC/77ºF) Low Range: ±0.5 mg/L or 3% of reading, whichever is greater; High Range: ±15 mg/L or 3% of reading, whichever is greater

Methods acid base titration, total acidity / strong acidity

Principle fixed endpoint titration: 8.30 pH (phenolphthalein) or 3.7 pH (methyl orange)

Pump Speed 10 mL/min

Stirring Speed 600 rpm

pH Range -2.0 to 16.0 pH; -2.00 to 16.00 pH

pH Resolution 0.1 pH / 0.01 pH

pH Accuracy (@25ºC/77ºF) ±0.01 pH

pH Calibration one, two or three-point calibration; four available buffers (4.01, 6.00, 8.30, 10.01)

Temperature Compensation manual or automatic

mV Range -2000.0 to 2000.0 mV

mV Resolution 0.1 mV

mV Accuracy (@25ºC/77ºF) ±1.0 mV

Temperature Range -20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K

Temperature Resolution 0.1°C; 0.1°F; 0.1K

Temperature Accuracy (@25ºC/77ºF) ±0.4°C; ±0.8°F; ±0.4 K

Logging Data up to 400 ( 200 titrations, 200 pH/mV readings)

Electrode HI1131B glass body, refillable, with BNC connector and 1 m (3.3’) cable (included)

Temperature Probe HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included)

Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing

Power Supply 12 VDC adapter (included)

Dimensions 235 x 200 x 150 mm (9.2 x 7.9 x 5.9”)

Weight 1.9 kg (67.0 oz.)

Ordering Information

HI84530 is supplied with the HI1131B pH electrode, HI7662-T temperature probe, HI84530-50 low range titrant, HI84530-51 high range titrant, HI84530-55 calibration standard, HI84530-60 hydrogen peroxide, 100 mL beakers (2), dosing pump valve, 5 mL syringe, 1 mL plastic pipette, tube set (aspiration tube with titrant bottle cap and dispensing tube with tip), stir bar, power adapter, instruction manual and quality certificate.