Your shopping cart is empty!
Availability: กรุณาสอบถามค่ะ
This instrument utilizes a four ring potentiometric probe with platinum sensor that offers greater versatility over typical amperometric designs. A potentiometric probe works on the principal of induction which eliminates the effects of polarization (a common problem of amperometric systems). Two outer rings apply an alternating voltage and induce a current loop in the solution while two inner rings measure the voltage drop induced by the current loop (which is dependent on the conductivity of the solution). By utilizing the 4-ring method, it is possible to measure very low or high conductivity levels (up to 200 mS/cm) without changing probes.
The temperature coefficient correction is settable between 0 and 2.5%/ºC.
Features:
Specifications:
Range | 0.0 to 199.9 µS/cm; 0 to 1999 µS/cm; 0.00 to 19.99mS/cm; 0.0 to 199.9 mS/cm |
|
Resolution | 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm |
|
Accuracy | ±1% F.S. (excluding probe error) | |
Calibration | manual, 1 point | |
Temperature Compensation | automatic, 0 to 50°C (32 to 122°F) with ß adjustable from 0 to 2.5%/°C | |
Probe | HI 76303, platinum 4-ring sensor, built-in temperature sensor, 1 m (3.3') cable (included) |
|
Power Supply | 12 Vdc adapter (included) | |
Environment | 0 to 50°C (32 to 122°F); RH max 95% | |
Dimensions | 240 x 182 x 74 mm (9.4 x 7.2 x 2.9") | |
Weight | 1.0 kg (2.2 lb.) |