Your shopping cart is empty!
Availability: กรุณาสอบถามค่ะ
1.) Step One - Add a sample to the included cuvette(s)
2.) Step Two - Insert sample into the Checker and zero it out
3.) Step Three - Remove sample and add reagent
4.) Step Four – Reinsert sample and click the button to measure your results or…
- Press and hold the button for 3 seconds for a time reaction reading
The HI 715 uses an adaptation of the ASTM Manual of water and Environmental Technology D1426-93, Nessler Method. The reaction between ammonia and reagents causes a yellow tint in the sample.
FEATURES/BENEFITS:
* Small palm-sized design
* Accurate from ±0.05ppm ±5% of reading
* Dedicated to a single parameter
* Uses a single AAA battery
* Auto Power Off after 10 minutes in inactivity
* Built-in timer for timed reaction readings
PRIMARY APPLICATIONS:
* Water Quality
* Education
* Aquariums
* Environmental
Order Information:
The HI 715 Checker®HC is supplied with 2 cuvettes and caps, reagent starter kit (1 bottle of HI 715A-0 and 1 bottle of HI 715B-0), 1 x 1.5V AAA battery, and an instruction manual.
Specification:
Range | 0.00 to 9.99 ppm NH3-N | |
Resolution | 0.01 ppm | |
Accuracy (@20°C/68°F) | ±0.05ppm ±5% of reading | |
Measurement Method | Colorimetric | |
Light Source | Photodiode @470 nm | |
Light Detector | Silicon photocell | |
Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-93, Nessler method. The reaction between ammonia and reagents causes a yellow tint in the sample. | |
Power Supply | Battery - 1 x 1.5V AAA | |
Power Auto Off | After 10 min. of inactivity | |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing | |
Dimensions | 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”) | |
Weight | 64 g (2.25 oz.) |