0
รวม 0.00 ฿
 x 

Your shopping cart is empty!

รวม 0.00 ฿

FREE DELIVERY !

3,000 บาทขึ้นไป*

Technical Support

ทดลองสินค้าหน้าร้าน บริการหลังการขาย

OPEN ( MON-SAT )

8.00-17.30(จ-ศ) 9.00-15.30(เสาร์)

Fluke 1507 Insulation Resistance Testers

Availability: กรุณาสอบถามค่ะ

The Fluke 1507 and 1503 Insulation Testers are compact, rugged, reliable, and easy to use. With their multiple test voltages, they are ideal for many troubleshooting, commissioning, and preventative maintenance applications.
สอบถามเกี่ยวกับสินค้า
Product Tags

เครื่องวัดความเป็นฉนวน FLUKE 1507

 

FLUKE 1507 Insulation Resistance Testers

Features:

  • Insulation test range:
    • 1507: 0.01 MΩ to 10 GΩ
    • 1503: 0.1 MΩ to 2000 MΩ
  • Insulation test voltages:
    • 1507: 50 V, 100 V, 250 V, 500 V, 1000 V
    • 1503: 500 V, 1000 V
  • Save both time and money with Automatic calculation of Polarization Index and Dielectric Absorption Ratio (only available on the1507)
  • Make repetitive tests simple and easy with the 1507’s Compare (Pass/Fail) function (only available on the 1507)
  • Repetitive or hard-to-reach testing is easy with the remote test probe
  • Live circuit detection prevents insulation test if voltage > 30 V is detected for added user protection
  • Auto-discharge of capacitive voltage for added user protection
  • AC/DC voltage: 0.1 V to 600 V
  • 200 mA Continuity
  • Resistance: 0.01 Ω to 20.00 kΩ
  • Save battery power with auto power off
  • Read measurements easily with large, backlit display
  • CAT IV 600 V overvoltage category rating for added user protection
  • Remote probe, test leads, probes and alligator clips included with each tester
  • Accepts optional Fluke TPAK magnetic hanging system to free your hands for other work
  • Four AA alkaline batteries (NEDA 15 A or IEC LR6) for at least 1000 insulation tests
AC/DC Voltage Measurement
Accuracy
Range  600.0 V
Resolution  0.1 V
50 Hz to 400 Hz ± (% of Rdg + Digits)  ± (2 % + 3)
Input impedance
   3 MΩ (nominal), < 100 pF
Common mode rejection ratio (1 kΩ unbalanced):
   > 60 dB at dc, 50 or 60 Hz
Overload Protection
   600 V rms or dc
Earth bond resistance measurement
Range/Resolution
20.00 Ω  0.01 Ω
200.0 Ω  0.1 Ω
2000 Ω  1.0 Ω
20.00 kΩ  0.01 kΩ
Accuracy
   ± (1.5 % + 3)
Overload Protection
   2 V rms or dc
Open Circuit Test Voltage
   > 4.0 V, < 8 V
Short circuit current
   > 200.0 mA
Insulation specifications
Measurement Range
1507:  0.01 MΩ to 10 GΩ
1503:  0.01 MΩ to 2000 MΩ
Test voltage
1507:  50 V, 100 V, 250 V, 500 V, 1000 V
1503:  500 V, 1000V
Test Voltage Accuracy
   + 20 %, - 0 %
Short circuit current
   1 mA nominal
Auto Discharge
   Discharge time < 0.5 second for C = 1 µF or less
Live circuit indicator
   Inhibit test if terminal voltage > 30 V prior to initialization of test
Maximum Capacitive Load
   Operable with up to 1 µF load
Measure Accuracy:
50 V:  ±(3 % + 5)
100 V:  ±(3 % + 5)
250 V:  ±(1.5 % + 5)
500 V:  ±(1.5 % + 5)
1000 V:  ±(1.5 % + 5) to 2000 MΩ, ±(10 % + 3) above 2000 MΩ
General Specifications
Maximum voltage applied to any terminal:
   600 V ac rms or dc
Storage Temperature
   -40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature
   -20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient
   0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F)
Relative Humidity
   0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
   0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
   0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration
   Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock
   1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor)
Electromagnetic compatibility:
   In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997)
Safety
   Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category IV 600 V (CAT IV)
Certifications:
   CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition
Batteries
   Four AA batteries (NEDA 15A or IEC LR6)
Battery Life
Insulation test use:  Tester can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.
Resistance measurements:  Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 Ω with a duty cycle of 5 seconds on and 25 seconds off.
Size
   5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight
   550 g (1.2 lb)
IP Rating
   IP40
Altitude
Operating:  2000 m CAT IV 600 V, 3000 m CAT III 600 V
Non-operating (storage):  12,000 m
Over-range capability:
   110 % of range
Included accessories:
   TL224 Test Leads, TP74 Test Probes, clips PN 1958654 (red) and PN 1958646 (black), holster and remote probe