1.This instrument is only suitable for single-phase 230V / 50Hz (with power supply voltage range
2.Wiring check: whether three LED indicators are correct, if two green indicators P-E and P-N are on, and red indicator is not on, then it is correct: the test can only be triggered if the wiring is correct.
3.Phase angle selection: you can choose to start the test from positive (0 °) or from half cycle negative (180 °);
4.“OL" is displayed for over range: when the test trip action time exceeds the maximum test time, the LCD will display "OL mS"
5.Automatic data holding: After test is completed, the displayed measurement results are held for a certain period of time;
6.AUTO RAMP test: The AUTO RAMP test can test tripping current and time simultaneously.
7.Shutdown alarm function: After 3 minutes of operation, the instrument will let out a shutdown beep alarm;
8.Built-in FUSE security protection;
9.Power supply: directly supplied by circuit under test (power supply 230V / 50Hz);
Specifications
1.1 Measurement range and measurement error (temperature:
23 + 5 ℃;Humidity:45%~75% RH; altitude≤2000m:)
Functions |
Operation voltage (AC) | Rated operating current(L△n) | Fault action time(MAX) | Accuracy | |
Action current | Action time | ||||
X1/2 | 230V(Error range:-15%~+10%)
Frequency:50Hz |
10/20/30/100/300/500mA | 1000ms | Error range:-10%~+10% | +0.6% rdg +8dgt |
X1 | 10/20/30/100/300mA | 1000ms | Error range:-10%~+10% | ||
500mA | 300ms | ||||
X5 | 10/20/30mA | 1000ms | |||
AUTO RAMP test | 10/20/30/100/300/500mA | (RAMPIncreasing step10%) L△n20%~110% |
Error range:-10%~+10% |