Buttons and LCD display
1.Main data display
2.Data hold
3.Data messuring
4.Laser light indication
5.Backlight indication
6.Low battery indication
7.Laser button: Press it (2 seconds) to open or close the laser light.
8.℃/℉ button: Press it (2 seconds) to convert to Celsius or Fahrenheit.
9.Back light: Press it (2 seconds) to open or close the backlight
Main Technical Indexes
1. Normal work conditions:
- Ambient temperature: 10℃~40℃
- Storage temperature: -10℃~40℃
- Relative humidity: ≤85%
- Power supply: two 1.5V AAA/LR03 battery
2. Basic size: 143mm×93mm×37mm (L×W×H)
3. Weight (net weight): 105g
4. LCD resolution (precision): 0.1℃/℉
5. Measuring range : 32℃~42.9℃ (89.6℉~109.2℉)
6. Power consumption: ≤50mw
7. Accuracy:
32~35.9℃:±0.3℃(0.6℉)
36~39℃:±0.2℃(0.6℉)
39.1~42.9℃:±0.3℃(0.6℉)
8. Measuring time: 0.5s
9. Measuring distance: 5-10cm
10. Automatic shutdown time: 15s
11. Safety design standard: Meeting European CE safety standard.
Operation method
1. The fixed emissivity of the thermometer is 0.95 by default;
2. In order to measure out exact value, after the battery is installed, put the meter under the temperature
for 25-30 minutes, then use it.
3. The head of the device with infrared on the forehead is measured with a distance of 5 to 10 cm from
the top of the device to the forehead and pull the trigger, about 3 to 4 seconds when the screen is stable
then leave your finger out from trigger, at this time you will hear a beep and at the same time, the measurement
results appear on the screen together with the hold function.
4. When the measured temperature is higher than 37.3c, there will be a beep alarm.
5. When the measured temperature exceeds 42.9c, the LCD will display "HI".
6. When the measured temperature is lower than 32c, the LCD will display "LO".
7. The information "HI" or "LO" appears. There are several situations for reference