Your shopping cart is empty!
Availability: กรุณาสอบถามค่ะ
Nickel is commonly utilized by the electroplating industry in processes utilizing stainless steel, cobalt or nickel alloys.
Nickel is also used in batteries, fuel cells and hydrogenation of vegetable oils in the food industry.
The HI 96726 measures the Nickel content in water and wastewater. This meter uses an exclusive positive-locking system to ensure that the cuvette is in the same place every time it is placed into the measurement cell.
The cuvette has a very important role because it is an optical element and thus requires particular attention. It is important that both, the measurement and the calibration (zeroing) cuvettes, are optically identical to provide the same measurement conditions.
Order Information:
HI 96726 is supplied with sample cuvettes with caps (2), 9V battery and instruction manual.
HI 96726C includes photometer, sample cuvettes with caps (2 ea.), 9V battery, scissors, cloth for wiping cuvettes, instrument quality certificate, instruction manual and rigid carrying case.
Range | 0.00 to 7.00 g/L | |
Resolution | 0.01 g/L | |
Accuracy | ±0.07 g/L ±4% @ 25°C | |
Light Source | Tungsten lamp | |
Light Detector | Silicon photocell with narrow band interference filter @ 575 nm | |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing | |
Power Supply | (1) 9V battery | |
Auto-off | After 10 minutes of non-use in measurement mode; after 1 hour of non-use in calibration mode; with last reading reminder. | |
Dimensions | 192 x 104 x 69 mm (7.6 x 4.1 x 2.7") | |
Weight | 360 g (12.7 oz.) | |
Method | Adaptation of the photometric method. |