Your shopping cart is empty!
Availability: กรุณาสอบถามค่ะ
The Kimo Kigaz 200 PRO features 3 interchangeable O2, NO and CO - H2 sensors along with a built in printer.
Feature
Display: 3.5 inch
Built in printer
CO sensor protection through solenoid valve
Led light on the handle
Single connector for draft and flue gas probe
Backlit 10 key pad with dome switch
Housing: ABS
PC interface: Bluetooth, USB and infrared
Power supply: 3.6V 4400 mA Li-Ion battery
Flue gas probe: 180 mm with a water tap
LIGAZ software: for creating database, downloading inspections, printing inspections
Dimensions: 331x112x86 mm
Includes
Differential pressure kit
Transport bag
Calibration certificate
Protective magnetic cover
Specification
Kigaz 200 PRO Technical Features |
|||||
GAS |
Ambient CO max |
CO flue gas |
Interchangeable (CH4 optional) |
Excess air Losses |
Efficiency > 100 percent |
PRESSURE |
Differential pressure measurement |
Draft measurement |
|
|
|
temp. |
Ambient temp. |
Flue gas temp. |
Delta temp. |
DHW temp. |
Dew point temp. |
OTHER FUNCTIONS |
15 programmed combustibles |
Adding 5 combustibles by the user |
Automatic measurements |
Opacity index |
- |
Combustible list - Sahara / Fos sur Mer Nat. Gas, Natural Gas (Groningen), North Sea / Russia Nat. Gas, LPG, Propane, Butane, Heavy Oil, Light Oil, Coke gas, Bituminous coal, Hard coal, Bio fuel 5 percent, Wood 20%, Wood chip 21%, Pellet 8% Specifications |
Kigaz 200 PRO Measuring Range |
|||||
Parameter |
Sensor |
Measuring range |
Resolution |
Accuracy* |
T90 response time |
O2 |
Electro-chemical |
0 percent - 21 percent |
0.1 percent vol. |
+/-0.2 percent vol. |
30 s |
CO |
Electro-chemical |
0 - 8000 ppm |
1 ppm |
0 - 200 ppm : +/-10 ppm |
30 s |
NO |
Electro-chemical |
0 - 5000 ppm |
1 ppm |
0 - 100 ppm : +/-5 ppm. |
30 s |
NOx |
Calculated ** |
0 - 5155 ppm |
1 ppm |
- |
- |
CO2 |
Calculated ** |
0 - 99 percent vol |
0.1 percent vol |
- |
- |
CH4 |
Semiconductor |
0 - 10000 ppm |
1 ppm |
+/-20 percent of full scale |
40 s |
Flue gas temp. |
K thermocouple |
-100 - +1250 deg. C |
0.1 deg. C |
+/-1 deg. C |
45 s |
Ambient temp. |
Internal NTC |
-20 - +120 deg. C |
0.1 deg. C |
0.1 deg. C |
45 s |
Ambient temp. |
Pt100 (1/3 Din ext. probe) |
-50 - +250 deg. C |
0.1 deg. C |
+/-0.3 percent of the measured value +/-0.25 deg. C |
30 s |
Dew point temp. |
Calculated ** |
0 - +99 deg. Ctd |
0.1 deg. C |
- |
- |
DHW temp. |
TcK (ext. probe) |
-200 - +1300 deg. C |
0.1 deg. C |
+/-1 deg. C |
- |
Differential pressure Draft |
Piezoelectric |
-20 000 - +20 000 Pa |
1 Pa |
-20 000 - -751 Pa : +/-(-0.5 percent of measured value +4.5 Pa) |
- |
Losses |
Calculated ** |
0 - 100 percent |
0.1 percent |
- |
- |
Flue gas velocity |
- |
0 - 99.9 m/s |
0.1 m/s |
- |
- |
Excess air λ |
Calculated ** |
1 - 9.99 |
0.01 |
- |
- |
Efficiency ηs |
Calculated ** |
0 - 100 percent |
0.1 percent |
- |
- |
Efficiency ηt(condensation) |
Calculated ** |
0 - 120 percent |
0.1 percent |
- |
- |
Opacity index |
ext. instrument |
0 - 9 |
- |
- |
- |
{tortags,2537,3}